Bath tubs care

Cleaning after each patient

After draining the water from the bath, open the valves of the nozzle rinsing for at least 1 minute. If there is insufficient water pressure in the system at the customer, it will be necessary to first open one periphery, rinse it for at least 1 minute and then open the other periphery after the previous one has been closed. Rinsing can be performed using hot, cold or mixed water. After the specific time has expired, close the valves. Then spray the bath cleanser PANTRA 05 onto the walls of the bath so that the entire surface of the bath is covered. Leave it for 2 minutes. After the specific time has expired, wash the bath using the handheld shower. After the bath has been washed, wipe it with a soft absorbent cloth. After the bath has been wiped dry, apply the disinfectant Bannox Hospital onto the surface of the bath, not on the control panel, using the sponge. Afterwards fill the bathtub for another patient. After the end of the shift, polish the chrome control elements of the bath on the control panel. Any possible hard water sediments can be removed using a cloth soaked in cleanser Pantra 05. After the sediments have been removed using a damp cloth, wipe the cleanser and polish the chrome elements with a dry cloth

Bath maintenance once in two weeks

After draining the water from the bath, unscrew the caps of hydromassaging and micromassaging nozzles, the caps of suction baskets and waste pipe. Remove the plastic balls of the nozzles from the large hydromassaging nozzles. After that spray the cleanser Pantra 05, or in case of severe dirt spray the cleanser Aktivit NK into the holes. Leave it for at least 5 minutes. Use the cleanser to spray the dismantled nozzles and caps.

Notice:

Clean the nozzles, caps in a separate container so that the small things do not fall into the waste pipe, of which the cap has also been removed. After that, fill the bath with water. The water level should be about 2 cm above the most upper nozzle. Pour 2 to 5l of Aktivit NK into the bath and start all systems of the bath for a period of 1 procedure, i.e. for 20 minutes.

Notice:

Before starting the bath, the regulation valves of the air intake in the bath nozzles must be closed to avoid large foaming in the bath. If this happens, turn off the bath using the handheld shower with cold water and reduce the level of foam in the bathtub. Wait until the foam settles, check the closing of the regulation valves of the air intake in the nozzles and start all systems of the bath again. After the specific time has expired, leave the cleanser in the bath for at least 10 minutes.  After draining the water from the bath, open the valves of the nozzle rinsing for at least 1 minute. If there is insufficient water pressure in the system, it will be necessary to first open one periphery, rinse it for at least 1 minute and then open the other periphery, after the previous one has been closed. Rinsing can be performed using hot, cold or mixed water. After the specific time has expired, close the valves and check if the openings of the nozzles are clean without any sediment; if there is sediment, spray the cleanser on it and mechanically remove it using the old toothbrush.

Wash the dismantled components using the clean water and install them back onto the bath. Spray the bath cleanser PANTRA 05 onto the walls of the bath so that the entire surface of the bath is covered. Leave it for 2 minutes. After the specific time has expired, wash the bath using the handheld shower. After the bath has been washed, wipe it with a soft absorbent cloth. After the bath has been wiped dry, apply the disinfectant Bannox Hospital onto the surface of the bath, not on the control panel, using the sponge. Polish the chrome control elements of the bath on the control panel, using the soft cloth. Any possible hard water sediments can be removed using a cloth soaked in cleanser Pantra 05. After the sediments have been removed using a damp cloth, wipe the cleanser and polish the chrome elements with a dry cloth.

Recommendation cleaning and disinfection of hydrotherapy MEDEXIM

Object of cleaning  Appliance  Spectrum of efficiency Dosing  Time of actuation in min. How to use  Interval of cleaning
    
  
Scottish squirts  AQUABELA     
Mechanical cleaning Pantra 05, GA 09 2dcl/1l 1 - 5 To lay on the jets and also to inside - the best created foam with the instrument for foaming, let to react, flush and wipe up to the dry state  Daily, at the end of working shift
WC TIP Wipe up the bath, let to react, flush with water
Baths at underwater massages/whirling machines    
Fumigation 5P plus A, B 1dcl/1l 1-5 Wipe up the bath, let to react, flush with water After each patient
GA05,Banox hospital A,B,T,V,(C) 2dcl/1l 1-5
Mechanical cleaning Pantra 05, GA 09 2 dcl/1l 1-5 To splash with the instrument for foaming, let to react, flush and wipe up to dry state Daily, at the end of the working shift
Sanitarian cleaning GA NK, Aktivit NK 2-5l/300l 30-60 To dilute to the bath, switch on the pump, let to react alternately to the massage jets and to the massage hose (as far as the bath is equipped with the hose)   After 100 procedes, at the further after two weeks
WC TIP Wipe up the bath, let to react, flush with water
Bubble grates
Fumigation 5P plus A, B 1dcl/1l 1-5 Wipe up the bath, let to react, flush with water After each patient
GA05,Banox hospital A,B,T,V,(C) 2dcl/1l 1-5
Mechanical cleaning Pantra 05, GA 09 2dcl/1l 1-5 To splash with the instrument for foaming, let to react, flush and wipe up to the dry state Daily, at the end of the working shift
Sanitarian cleaning GA NK, Aktivit NK 2-5l/300l 30 After mechanical cleaning to let submerged in the solution After 100 procedures, at the further after two weeks
WC TIP Wipe up the bath, let to react, flush with water
Balneologic bathes    
Fumigation 5P plus A, B 1dcl/1l 1-5 Wipe up the bath, let to react, flush with water After each patient
GA05,Banox hospital A,B,T,V,(C) 2dcl/1l 1-5
Mechanical cleaning Pantra 05, GA 09 2dcl/1l 1-5 To splash with the instrument for foaming, let to react, flush and wipe up to the dry state Daily, at the end of the working shift
WC TIP Wipe up the bath, let to react, flush with water

Useful advice

  • When cleaning the balneotherapeutic devices in the area of control  panel, it is necessary to avoid pouring of chemical cleaning agents to the buttons /both mechanical and electrical/ since there is a risk of their damage and loss of function. Treat them in a suitable considerable way, e.g. with a moist cloth.,

    

  • It is unallowable to rinse the control panel by water jet, the buttons might  bunged up and became inoperative,

    

  • Only the agents recommended by or consulted with the manufacturer may be used for cleaning, the manufacturer shall bear no responsibility for damage caused by the incorrect use of cleaning agents,

    

  • In no case the lime scale and impurity sediments may not removed in a mechanic way,
  • When cleaning the devices it is not allowed to rinse the individual jets using water pressure, there is a risk of an irreversible damage to some motors, rinse that is the part of the device is intended for cleaning the distribution system ,
  • Never use decolorants, acetone, toluene, thinners on a synthetic base and other organic solvents for  cleaning. Should you use the unsuitable cleaner, the damage to the structure shall occur, which shall not be covered by the warranty.
  • We do not recommend using liquid cleaners comprising abrasives (sand) that cause the scratching of the acrylate surface. In no case use  scourers, metal sponges, knives, steel wool or any aids that may damage the tub surface.  
  • Avoid the contact of tub surface with solvents, lacquers and thinners, including nail polishes or nail polish remover since they might damage their surface, Hot objects – cigarette, aromatic candle , etc. are dangerous as well.
  • The oil or greasing traces may be removed using denaturized spirit. In no case use strong solvents, such as acetone, paint thinners, petrol, and ammonia or chloride preparations.